Philips Defibrillator / Monitor Model DFM 100 Efficia
The Efficia DFM100 is an affordable solution with the ability to adapt to changes in your clinical environment, resource needs, or critical care guidelines.
Model name/Number | DFM 100 Efficia |
Type | ALS defibrillators |
Brand | Philips |
Usage/Application | Hospitals ICU |
Operation mode | Manual |
Battery type | Rechargeable lithium ion battery |
Display type | Color TFT LCD |
ECG monitoring | Yes |
a) Physical characteristics
Filtration – Rechargeable lithium ion battery; AC power using a protectively grounded outlet.
b) General
Approximate dimensions – 23.5 cm (H) x 29 cm (W) x 20.5 cm (D); 9.25 in (H) x 11.4 in (W) x 8 in (D)
Standard operator position – Within one meter (3 feet) of the device
Approximate weight (without battery) – 5.66 kg; 12.5 lbs
Power – Rechargeable lithium ion battery; AC power using a protectively grounded outlet
c) Defibrillator
Waveform – Approximately 7 in (17.8 cm) diagonal viewing area
Shock delivery – Via multifunction electrode pads or paddles
Shock series – Configurable energy escalation in a series leads off sensing and PCI
Charge time – Less than 5 seconds to the recommended adult energy level (150 joules) with a new, fully-charged battery installed.
Less than 6 seconds to the selected energy level (up to 200 joules) with a new, fully charged battery installed.
Less than 15 seconds to the selected energy level while connected to AC power only.
The device powers on in manual defibrillation mode ready to deliver shock in less than – 15 seconds with a new, fully-charged battery.
The device powers on in AED mode ready to deliver shock in less than – 24 seconds with a new, fully-charged battery even after 15 discharges of maximum energy.
Sensing for pads/paddles – Apply 500 nA rms (571 Hz); 200 uA rms (32 KHz)
Patient impedance range – Minimum: 25 ohm (external defibrillation); 15 ohm (internal defibrillation) – Maximum: 250 ohm. Actual functional range may exceed these values
d) Display
Size – Approximately 7 in (17.8 cm) diagonal viewing area
Type – Color TFT LCD
Resolution – 800 x 480 pixels (VGA) with 32 brightness levels per color
Wave viewing time – 6.5 sec ± 10%
Sweep speed – 25 mm/s ± 10% nominal (stationary trace; sweeping erase bar) for ECG and SpO2; capnogram wave is 6.25 mm/s ± 10%